Field Metrology Well, -25 °C to 150 °C, w/9142-INSX, w/Process Electronics
Temperature Range at 23 °C: -25 °C to 150 °C (-13 °F to 302 °F)
Display Accuracy: ± 0.2 °C Full Range
Stability: ± 0.01 °C Full Range
Axial Uniformity at 40 mm (1.6 in): ± 0.05 °C Full Range
Radial Uniformity: ± 0.01 °C Full Range
Loading Effect (with a 6.35 mm reference probe and three 6.35 mm probes): ± 0.006 °C Full Range
Hysteresis: 0.025 °C
Operating Conditions: 0 °C to 50 °C, 0 % to 90 % RH (non-condensing)
Environmental Conditions: 13 °C to 33 °C
Immersion (Well) Depth: 150 mm (5.9 in)
Insert OD: 30 mm (1.18 in)
Heating Time: 16 min: 23 °C to 140 °C, 23 min: 23 °C to 150 °C, 25 min: -25 °C to 150 °C
Cooling Time: 15 min: 23 °C to -25 °C, 25 min: 150 °C to -23 °C
Resolution: 0.01 °
Display: LCD, °C or °F user-selectable
Size (H x W x D): 290 mm x 185 mm x 295 mm (11.4 x 7.3 x 11.6 in)
Weight: 8.16 kg (18 lb)
Process version provides a built-in two-channel thermometer readout that measures PRTs, RTDs, thermocouples, and 4-20 mA transmitters which includes the 24 V loop supply to power the transmitter.