Fluke Networks CFP-Q-ADD-R CertiFiber Pro 2 Quad Add On Kit with Remote
CertiFiber Pro 2 Quad Add On Kit with Remote
Order #: CFP2-Q-ADD-R
Mfg #: 4955957
$13,942.99
Fluke Networks CFP-Q-ADD-R CertiFiber Pro 2 Quad Add On Kit with Remote
CertiFiber Pro 2 Quad Add On Kit with Remote
Order #: CFP2-Q-ADD-R
Mfg #: 4955957
$13,942.99
The Fluke Networks CFP-Q-ADD-R is a CertiFiber Pro Quad Add On Kit W/Remote. This item should be ordered when you need a CertiFiber Pro Quad (SingleMode and MultiMode) optical fiber certification add-on kit with a remote unit, modules and accessories, WITHOUT a mainframe (upgrading from OptiFiber Pro). The Fluke Networks Versiv family of cable testers offer modular solutions for every copper and fiber cable testing application. Check out our CertiFiber Pro product page to view all of your ordering options.
Availability: Ships in 3-4 weeks
Product Highlights
- Fastest time to certify two fibers at two wavelengths in three seconds.
- Encircled Flux compliant as required by ANSI/TIA and ISO/IEC
- Merged Tier 1 (Basic) / Tier 2 (Extended) testing and reporting when paired with OptiFiber Pro
- ProjX management system ensures all jobs are completed correctly and efficiently
- Provides automatic Pass/Fail analysis to industry standards or custom test limits
- Double ended Pass/Fail certification of fiber optic connector end faces
- Graphical indication of problem areas on fiber end faces due to contamination, pits, chips and scratches
- Versiv modular design supports copper certification, fiber optic loss, OTDR testing and fiber end-face inspection
- Taptive user interface provides simple, animated guidance to eliminate incorrect reference setup and negative losserrors.
- Convenient quad module supports both multimode and singlemode loss testing
- Extended singlemode distance range to 130km
- Built in Visual Fault Locator
- Analyze test results and create professional test reports using LinkWare Management Software
- Compatible with Linkware Live. Linkware Live enables to easily track job progress, get real-time access to test results to quickly fix problems in the field, and easily transfer and consolidate test results from the tester to LinkWare PC Cable Test Management Software.
Each Unit Includes
- 1 Versiv Remote
- (2) CertiFiber Pro Quad OLTS modules
- SC/LC EF-compliant MM TRC Kit 50 µm
- SC/LC SM TRC Kit
- TRC Carrying Case
- (2) LC/LC Simplex Adaptors
- (2) HandStraps
- (2) Shoulder Straps
- Large Carry Case
- Versiv Open Source Software CD
- AC Chargers
- Statement of Calibration and Getting Started Guide.
Specifications
Power Meter Specifications | |
Input Connector | Interchangeable connector adapter (LC standard, SC, ST and FC optional) |
Detector Type | InGaAs |
Wavelengths | 850 nm, 1300 nm, 1310 nm, 1550 nm |
Power Measurement Range | 0 dBm to -65 dBm (850 nm) 0 dBm to -70 dBm (all other wavelengths) |
Power Measurement Uncertainty1,2 | < +/- 5% +/- 32 pW |
Measurement Linearity3 | < ± 0.1 dB |
Re-calibration Period | 1 year |
1. +/- 100 pW at 850 nm 2. Under the following conditions: Power level 100 _W (-10 dBm), continuous wave (CW) for absolute power at 850 nm and 1310 nm. Divergent beam, NA = 0.20 for 50/125 _m and NA = 0.14 for 9/125 _m. Ambient temperature 23° ± 1°C. SC/UPC connector with ceramic ferrule. After a 5 minute warm-up. Traceable to NIST. 3. -3 dBm to -55 dBm at 850 nm and 1310 nm. Ambient temperature 23° ± 1°C. After a 5 minute warm-up. |
Loss/Length Specifications | ||
Specification | CertiFiber Pro Multimode Modules | CertiFiber Pro Singlemode Modules |
Testing Speeds (excludes referencing times) | Smart Remote mode: < 3 seconds (2 wavelengths, one direction, auto wavelength detection) Far End Source mode: ≤ 2 seconds (2 wavelengths, one direction auto wavelength detection) Loopback mode: ≤ 2 seconds (2 wavelengths, one direction auto wavelength detection) | |
Input/Output Connectors | Interchangeable connector adapter (LC standard, SC, ST and FC optional) | |
Launch Condition1,2 | Encircled Flux compliant to TIA-526-14-B, ISO/IEC 14763-3 and IEC 61280-4-1 | |
Fiber Types Tested | 50/125 µm or 62.5/125 µm | singlemode |
Source Type and Wavelengths | LED source 850 nm ± 30 nm 1300 nm ± 20 nm | Fabry-Perot laser diode 1310 nm ± 20 nm 1550 nm ± 30 nm |
Maximum Length Measurement | 12 km | 130 km |
Length Measurement Accuracy | ± 1.5 m plus ± 1 % of length | ± 1.5 m plus ± 1 % of length |
Output Power (Nominal) | ≥ -24 dBm with EF-TRC | ≥ -4 dBm |
Output Power Stability3 | ± 0.05 dB over 8 hours ± 0.03 dB over 15 minutes | ± 0.1 dB over 8 hours ± 0.08 dB over 15 minutes |
1. At the output of the EF-TRC 2. Variations between EF measurement equipment may occur but EF compliance can be expected with a 95% confidence factor 3. Relative to power level after 15 minute warm up – at constant temperature |
Loss/Length Specifications (continued) | ||
Source Wavelengths | 850 nm, 1300 nm, 1310 nm, 1550 nm | |
Power Measurement Range | 0 dBm to -65 dBm (850 nm) 0 dBm to -70 dBm (all other wavelengths) | |
Power Measurement Uncertainty1, 2 | < +/- 5% +/- 32 pW | |
Measurement Linearity3 | < ± 0.1 dB | |
Re-calibration Period | 1 year | |
1. +/- 100 pW at 850 nm 2. Under the following conditions: Power level 100 _W (-10 dBm), continuous wave (CW) for absolute power at 850 nm and 1310 nm. Divergent beam, NA = 0.20 for 50/125 _m and NA = 0.14 for 9/125 _m. Ambient temperature 23° ± 1°C. SC/UPC connector with ceramic ferrule. After a 5 minute warm-up. Traceable to NIST. 3. -3 dBm to -55 dBm at 850 nm and 1310 nm. Ambient temperature 23° ± 1°C. After a 5-minute warm-up. |
Selected Specifications | |
Visual Fault Locator (VFL) | |
Title | Requirement |
Output Power | >-5 dBm <0 dBm SMF-28 singlemode fiber Continuous wave SC/UPC connector |
Operating Wavelength | 650 nm nominal |
Output Modes | Continuous wave Pulsed mode (2 Hz to 3 Hz blink frequency) |
Connector Adapter | 2.5 mm universal |
Laser Safety | Class II CDRH Over operating temperature range |
Environmental Specifications | |
Title | Requirement |
Operating Temperature | -18° C to 45° C |
Storage Temperature | -30° C to +60° C |
Operating Relative Humidity % RH without condensation | 0 % to 90 %, 32 °F to 95 °F (0° C to 35° C) 0 % to 70 %, 95 °F to 113 °F (35° C to 45° C) |
Vibration | Random, 2 g, 5 Hz-500 Hz |
Shock | 1 m drop test with and without module and adapters |
Safety | CSA C22.2 No. 1010.1: 1992 EN 61010-1 1st Edition + Amendments 1, 2 |
Pollution Degree | 2 |
Altitude | Operating: 4,000 m; storage: 12,000 m |
EMC | EN 61326–1 |
All products in this series
-
Fluke Networks CFP-MM MOD CertiFiber Pro Multimode Replacement ModuleCertiFiber Pro Multimode Replacement Module$4,735.99
-
Fluke Networks CFP-100-QI/GLD CertiFiber Pro 2 Quad OLTS, Inspection, Wi-Fi, 1 Year SupportCertiFiber Pro 2 Quad OLTS, Inspection, Wi-Fi, 1 Yr Support$23,088.99
-
Fluke Networks CFP-100-Q/GLD CertiFiber Pro 2 Quad OLTS with Wi-Fi and 1 Year Gold SupportCertiFiber Pro 2 Quad OLTS with Wi-Fi and 1 Year Gold Support$17,522.99
-
Fluke Networks CFP-100-S-W CertiFiber Pro 2 Singlemode OLTS with Wi-FiCertiFiber Pro 2 Singlemode OLTS with Wi-Fi$11,583.99
-
Fluke Networks CFP-MM-ADD CertiFiber Pro Singlemode Modules Add on KitCertiFiber Pro Singlemode Modules Add on Kit$6,802.99
-
Fluke Networks CFP-SM-ADD CertiFiber Pro Sm Modules Add On KitCertiFiber Pro Sm Modules Add On Kit$8,365.99
-
Fluke Networks CFP-QUAD MOD CertiFiber Pro Quad Replacement ModuleCertiFiber Pro Quad Replacement Module$7,502.99
-
Fluke Networks CFP-SM MOD CertiFiber Pro Singlemode Module OLTS ReplacementCertiFiber Pro Singlemode Module OLTS Replacement$5,940.99
-
Fluke Networks CFP-100-M CertiFiber PRO 2 Multimode OLTS with Wi-FiCertiFiber PRO 2 Multimode OLTS with Wi-Fi$10,122.99
-
Fluke Networks CFP-100-QI CertiFiber Pro 2 Quad OLTS with Double Ended Inspection & Wi-FiCertiFiber Pro 2 Quad OLTS with Double Ended Inspection & Wi-F$20,534.99