Product Highlights

The Test Equity 123H Temperature / Humidty Chamber provides high-end chamber performance in a smaller, upright package. The 123H comes standard with the newer F4T Touch Screen Controller. It's very easy and intuitive to operate. It includes RS-232 and Ethernet interfaces. GPIB is available as an option.

Features and specifications of the Test Equity 123H Temperature and Humidty Chamber include:

  • Temperature, Lowest: -68C
  • Temperature, Highest: +175C
  • LN2 Boost Cooling: No
  • Interior Volume: 2.3 Cu Ft
  • Input Voltage: 120V, 60Hz, 1 PH
  • 2.3 Cu Ft Workspace (18" W x 16.5" H x 13.5" D)
  • Humidity 10% to 95% RH. See specifications page for limitations.
  • GN2 Purge System Included
  • Four Event Relays Included
  • Your choice of the NEW F4T Touch Screen Controller or the Original F4 Controller
  • High/Low Limit Control and Alarm
  • Viewing Window & Interior Light
  • 4" Access Ports on Left & Right Side
  • Non-CFC Cascade Refrigeration

The TestEquity 123H also features:

High/Low Temperature Limit Controller and Alarm

  • Provides independent protection against excess temperatures. Both high and low limits can be set. The upper LED readout indicates the actual chamber temperature while the lower readout indicates the status The limit controller will shut down the chamber and trigger the audible alarm in the event of an out of limit condition. A set of uncommitted contacts are available to provide a safety power interlock for your test sample.

Accurate Humidity Measurement

  • A fast responding Vaisala capacitive-type sensor eliminates the need for unreliable wet-wicks.

GN2 Purge Capability Standard

  • For low humdity conditions below 6C Dewpoint (requires customer-supplied gaseous nitrogen supply).

Optional Dry Air Purge Capability - Field Upgradeable

  • Optional Dry Air Purge is easy to add in the field (requires customer-supplied compressed air supply).

Event Relays Standard

  • Four uncommitted solid-state relays are provided to switch the power to remote AC devices via the F4 Temperature/Humidity Controller.

120 Volt Input

  • The chamber draws 18 Amps maximum at 120 Volts. You can plug it into any standard 20 Amp wall receptacle.

Viewing Window

  • Heated multi-pane window eliminates fog and condensation without a manual wiper.<.li>

    Two 4" Access Ports

    • Lets you attach wires and sensors to your test sample through both the left and right side of the chamber.<.li>

      Adjustable Shelf

      • A heavy-duty stainless steel wire shelf ensures proper airflow around your test sample for uniform temperature distribution.<.li>

        Casters

        • Rugged casters permit easy movement.<.li>

          Accurate Humidity Measurement

          • A fast responding capacitive-type sensor eliminates the need for unreliable wet-wicks.<.li>

            Reliable Vapor Generator

            • Unlike "atomizing" systems, the heated vapor generator has no pump to fail or orifice to clog. The stainless-steel vapor generator assembly is double-protected against low-water and over-temperature conditions.<.li>

              No-Sweat Cabinet

              • We use a combination of structural foam and fiberglass blanket insulation for a higher combined R-factor than just fiberglass alone. The door sealing surface is kept warm by circulating excess heat from the refrigeration system. The result is a cabinet that does not sweat due to condensation during low temperature operation.<.li>

                Water Recirculation System

                • The optional water recirculation system provides a reliable source of demineralized water for the humidity system, using ordinary tap or pre-filtered water. Condensate from the chamber is returned back to the water tank. The recirculation pump operates only when needed to fill the chamber's humidifier to the correct level. Perfect for installations where a water line and drain are not available. <.li>

                  Reliability & Safety That's Designed In

                  • Non-CFC refrigerants are used in a cascade (two compressors) configuration. Expansion valves provide optimum performance throughout the operating range. Pressure regulators keep the operating pressures within safe limits regardless of load conditions. Liquid injection ensures cool compressor operation during high-temperature pull down for long life. Sequential starting of each compressor reduces the current demand on start-up. Each compressor is protected by high/low pressure switches. Care has been taken to use a minimum number of fittings, joints and welds to reduce the possibility of refrigerant leaks.<.li>
                  • All electrical control components are UL approved. The nichrome air-heater has a low watt-density and even heat distribution for reliable performance and fast response. A fusible link provides fail-safe protection against thermal runaways, in addition to the microprocessor-based high/low limit controller.<.li>
                  • The side and rear panels are removable to permit complete access to the refrigeration and humidity system for ease of service and preventive maintenance.<.li>