Your company account is blocked and you cannot place orders. If you have questions, please contact your company administrator.
Yokogawa 701265 for Rent, High Precision Temperature Input Module
High Precision Temperature Input Module

Order #: 701265-RENT3

Mfg #: 701265

Questions about availability or need help? Contact Us:

(800) 633-0561
Call for price
Call for price

Yokogawa 701265 for Rent, High Precision Temperature Input Module

High Precision Temperature Input Module

Order #: 701265-RENT3

Mfg #: 701265

Call for price

Request a Rental Quote









Have a Question? Speak to a Transcat Rental Expert: 800.264.4059

Call for price

Product Highlights

The Yokogawa 701265 High Precision Temperature Input Module provides 2 channels of temperature (thermocouple) or voltage (switchable) measurement.

Specifications of the Yokogawa 701265 High Precision Temperature Input Module include:

  • Function:  Temperature (thermocouple) or voltage measurement (switchable)
  • Effective measurement range:  Voltage mesurement: twice of setting range
  • Number of input channels:  2
  • Input coupling:
    • TC, DC, and GND
    • TC:  Temperature (thermocouple) measurement 
    • DC:  Voltage measurement (DC coupling) 
  • Voltage measurement maximum sample rate:  100 kS/s 
  • Temperature measurement data update rate:  500 Hz
    • Voltage:
      • rms:  1.5 mV 
      • peak-to-peak:  10 mV 
  • Measurement range/accuracy:
    • Voltage measurement:
      • Voltage sensitivity:  1 mV to 100 V range (1-2-5 steps) 
      • Voltage accuracy:  +/-(0.08% or range + 2u V) 
  • Frequency characteristics (-3 dB point when a sine wave of amplitude 60% of range is input): 
    • Voltage measurement: DC to 100 Hz
    • Temperature measurement:  DC to 100 Hz
  • Maximum input voltage (at a frequency of 1 kHz or less):
    • Temperature input:  42 V (DC + ACpeak)
    • Voltage input:  42 V (DC + ACpeak)
  • Maximum allowable common mode voltage (at a frequency of 1 kHz or less):
    • Temperature input:  42 V (DC + ACpeak)(CAT l and CAT ll, 30 Vrms)
    • Voltage input:  42 V (DC + ACpeak)(CAT l and CAT ll, 30 V)
  • Vertical resolution:
    • Voltage measurement during voltage input:  2400 LSB/div
    • Temperature measurement when measuring temperature:  0.1C
  • -3 dB point when AC coupled low frequency attenuation point:  Voltage measurement 0.5 Hz or less
  • Input connector:  Binding post
  • Input impedance:  Approximately 1 MOhm
  • Common mode rejection ratio:
    • Voltage measurement:  80 dB (50/60 Hz) or more typical
    • Temperature measurement:  120 dB or more (50/60 Hz, with 2-Hz filter ON, signal source resistance of 500 ohm or less) typical
  • Residual noise level (input section shorted):  
    • Voltage measurement: +/-4 u V or +/-0.1% of range, whichever is greater (typical)
  • A/D conversion resolution:
    • Voltage measurement:  16 bits (24000 LSB/range)
  • Temperature coefficient:
    • Voltage measurement:
      • Zero point:  +/-(0.01% of range)/ C (typical)
      • Gain:  +/-(0.02% of range)/ C (typical)
  • Bandwidth limit (digital filter):  Select from OFF, 30 Hz, 8 Hz, and 2 Hz
  • Input bias current:  20 nA or less